Potrebujeme váš súhlas na využitie jednotlivých dát, aby sa vám okrem iného mohli ukazovať informácie týkajúce sa vašich záujmov. Súhlas udelíte kliknutím na tlačidlo „OK“.
Standard Practice for Pulse Counting System Dead Time Determination by Measuring Isotopic Ratios with SIMS
Automaticky preložený názov:
Štandardné praktiky pre Pulse systém počítania nečinnej doby Stanovenie meraním izotopové pomery s SIMS
NORMA vydaná dňa 1.6.2010
Označenie normy: ASTM E2426-10
Poznámka: NEPLATNÁ
Dátum vydania normy: 1.6.2010
Kód tovaru: NS-45280
Počet strán: 3
Približná hmotnosť: 9 g (0.02 libier)
Krajina: Americká technická norma
Kategória: Technické normy ASTM
Keywords:
dead time, isotopic ratios, pulse counting, SIMS, Isotopic analysis, Pulse testing, SIMS (secondary ion mass spectrometry), Surface analysis, ICS Number Code 71.040.50 (Physicochemical methods of analysis)
Significance and Use | ||||
Electron multipliers are commonly used in pulse-counting mode to detect ions from magnetic sector mass spectrometers. The electronics used to amplify, detect and count pulses from the electron multipliers always have a characteristic time interval after the detection of a pulse, during which no other pulses can be counted. This characteristic time interval is known as the “dead time.” The dead time has the effect of reducing the measured count rate compared with the “true” count rate. In order to measure count rates accurately over the entire dynamic range of a pulse counting detector, such as an electron multiplier, the dead time of the entire pulse counting system must be well known. Accurate count rate measurement forms the basis of isotopic ratio measurements as well as elemental abundance determinations. The procedure described herein has been successfully used to determine the dead time of counting systems on SIMS instruments. The accurate determination of the dead time by this method has been a key component of precision isotopic ratio measurements made by SIMS. |
||||
1. Scope | ||||
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for determining the dead time of the pulse-counting detection systems on the instrument. This practice also allows the analyst to determine whether the apparent dead time is independent of count rate. 1.2 This practice is applicable to most types of mass spectrometers that have pulse-counting detectors. 1.3 This practice does not describe methods for precise or accurate isotopic ratio measurements. 1.4 This practice does not describe methods for the proper operation of pulse counting systems and detectors for mass spectrometry. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||
2. Referenced Documents | ||||
|
Historická
1.6.2011
Historická
1.4.2012
Historická
1.10.2010
Historická
1.1.2014
Historická
1.10.2009
Historická
1.6.2005
Poskytovanie aktuálnych informácií o legislatívnych predpisoch vyhlásených v Zbierke zákonov od roku 1945.
Aktualizácia 2x v mesiaci !
Chcete vedieť viac informácii ? Pozrite sa na túto stránku.
Posledná aktualizácia: 2024-11-22 (Počet položiek: 2 206 568)
© Copyright 2024 NORMSERVIS s.r.o.