Standard Test Methods for Determining Radiopacity for Medical Use
Automaticky preložený názov:
Štandardné testovacie metódy pre stanovenie rádioopacitu pre lekárske použitie
NORMA vydaná dňa 1.3.2007
Označenie normy: ASTM F640-07
Poznámka: NEPLATNÁ
Dátum vydania normy: 1.3.2007
Kód tovaru: NS-55981
Počet strán: 4
Približná hmotnosť: 12 g (0.03 libier)
Krajina: Americká technická norma
Kategória: Technické normy ASTM
Keywords:
implants, radiopacity, surgical applications, surgical devices, surgical implants, test methods, X-ray, ICS Number Code 11.120.01 (Pharmaceutics in general), 83.080.01 (Plastics in general)
Significance and Use | ||||||||||
These methods are intended to determine whether a material, product, or part of a product has the degree of radiopacity desired for its application as a medical device in the human body. These methods allow for both qualitative and quantitative evaluation in different comparative situations. |
||||||||||
1. Scope | ||||||||||
1.1 These test methods cover the determination of the radiopacity of materials and products utilizing X-ray based techniques, including fluoroscopy, angiography, CT (computed tomography) and DEXA, also known as DXA, (dual energy X-ray absorptiometry). The results of these measurements are an indication of the likelihood of locating the product within the human body. 1.2 Types of Tests - There are three methods of tests described, differing in the method of determining radiopacity. 1.2.1 Method A - Radiopacity is (1) qualitatively determined by viewing image(s) of a test sample and the image background, with or without the use of a body mimic, or ( 2) quantitatively determined as a specific difference in optical density or pixel intensity between the image of a test sample and the image background, with or without the use of a body mimic. 1.2.2 Method B - Radiopacity is determined by (1) qualitatively comparing image(s) of a test sample and a user-defined standard without the use of a body mimic, or (2) quantitatively determining the specific difference in optical density or pixel intensity between the image of a test sample and the image of a user-defined standard without the use of a body mimic. 1.2.3 Method C - Radiopacity is determined by (1) qualitatively comparing image(s) of a test sample and a user-defined standard with the use of body mimic or (2) quantitatively determining the specific difference in optical density or pixel intensity between the image of a test sample and the image of a user-defined standard with the use of a body mimic. 1.3 The values stated in SI units are to be regarded as the standard. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||||||||
2. Referenced Documents | ||||||||||
|
Chcete mať istotu, že používate len platné technické normy?
Ponúkame Vám riešenie, ktoré Vám zaistí mesačný prehľad o aktuálnosti noriem, ktoré používate.
Chcete vedieť viac informácií ? Pozrite sa na túto stránku.
Posledná aktualizácia: 2024-05-19 (Počet položiek: 2 902 277)
© Copyright 2024 NORMSERVIS s.r.o.