Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging (Includes all amendments And changes 12/31/2010).
Automaticky preložený názov:
Štandardná skúšobná metóda pre meranie rádiové Indukované Kúrenie Near Pasívne implantátov Počas magnetickou rezonanciou
NORMA vydaná dňa 10.11.2002
Označenie normy: ASTM F2182-02a
Poznámka: NEPLATNÁ
Dátum vydania normy: 10.11.2002
Kód tovaru: NS-53107
Počet strán: 8
Približná hmotnosť: 24 g (0.05 libier)
Krajina: Americká technická norma
Kategória: Technické normy ASTM
Keywords:
implant, MRI (magnetic resonance imaging), MR safety, RF (Radio Frequency) heating, ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)
Significance and Use | ||||||||||||
This test method describes a test procedure for evaluating the RF-induced temperature rise in MRI in the vicinity of an implanted medical device. The actual temperature rise in the patient will depend on a variety of factors beyond the SAR and time of RF application. The conditions and results of the testing should be included in the device labeling so that the attending physician can make the decision of whether to allow the patient with the implant to undergo an MRI procedure. |
||||||||||||
1. Scope | ||||||||||||
1.1 This test method covers measurement of Radio Frequency (RF) induced heating near a passive medical implant and its surroundings during Magnetic Resonance Imaging (MRI). 1.2 This test method is one of those required to determine if the presence of a passive implant may cause injury to the person with the implant during an MRI procedure. Other safety issues that should be addressed include magnetically induced displacement force and torque. 1.3 The amount of RF-induced temperature rise for a given specific absorption rate (SAR) will depend on the RF frequency, which is proportional to the static magnetic field strength. Because of possible additional heating, particularly when device dimensions exceed a quarter wavelength, conclusions from measurements made at one frequency may not apply to other frequencies. 1.4 This test method assumes that testing is done on devices that will be entirely inside the body. 1.5 This test method applies to whole body magnetic resonance equipment, as defined in section 2.2.103 of the IEC Standard 60601-2-33 with a whole body RF transmit coil as defined in section 2.2.100. The RF coil is assumed to have quadrature excitation. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||||||||||
2. Referenced Documents | ||||||||||||
|
Chcete mať istotu o platnosti využívaných predpisov?
Ponúkame Vám riešenie, aby ste mohli používať stále platné (aktuálne) legislatívne predpisy
Chcete vedieť viac informácií ? Pozrite sa na túto stránku.
Posledná aktualizácia: 2025-09-28 (Počet položiek: 2 235 583)
© Copyright 2025 NORMSERVIS s.r.o.