Potrebujeme váš súhlas na využitie jednotlivých dát, aby sa vám okrem iného mohli ukazovať informácie týkajúce sa vašich záujmov. Súhlas udelíte kliknutím na tlačidlo „OK“.
Standard Test Methods for Leaks Using the Mass Spectrometer Leak Detector in the Detector Probe Mode
Automaticky preložený názov:
Štandardné metódy skúšok pre tesnosť Hmotnostný spektrometer detektor úniku v detektore režime Probe
NORMA vydaná dňa 1.12.2006
Označenie normy: ASTM E499-95(2006)
Poznámka: NEPLATNÁ
Dátum vydania normy: 1.12.2006
Kód tovaru: NS-46992
Počet strán: 6
Približná hmotnosť: 18 g (0.04 libier)
Krajina: Americká technická norma
Kategória: Technické normy ASTM
Keywords:
bell jar leak test, bomb mass spectrometer leak test, helium leak test, helium leak testing, leak testing, mass spectrometer leak testing, sealed object mass spectrometer leak test, ICS Number Code 71.060.01 (Inorganic chemicals in general)
Significance and Use | ||||
Test Method A is frequently used to test large systems and complex piping installations that can be filled with a trace gas. Helium is normally used. The test method is used to locate leaks but cannot be used to quantify except for approximation. Care must be taken to provide sufficient ventilation to prevent increasing the helium background at the test site. Results are limited by the helium background and the percentage of the leaking trace gas captured by the probe. Test Method B is used to increase the concentration of trace gas coming through the leak by capturing it within an enclosure until the signal above the helium background can be detected. By introducing a calibrated leak into the same volume for a recorded time interval, leak rates can be measured. |
||||
1. Scope | ||||
1.1 These test methods cover procedures for testing and locating the sources of gas leaking at the rate of 4.5 x 10 13 mol/s (1 x 108 Std cm 3/s) or greater. The test may be conducted on any device or component across which a pressure differential of helium or other suitable tracer gas may be created, and on which the effluent side of the leak to be tested is accessible for probing with the mass spectrometer sampling probe. 1.2 Two test methods are described: 1.2.1 Test Method A - Direct probing, and 1.2.2 Test Method B - Accumulation. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||
2. Referenced Documents | ||||
|
Chcete mať istotu, že používate len platné technické normy?
Ponúkame Vám riešenie, ktoré Vám zaistí mesačný prehľad o aktuálnosti noriem, ktoré používate.
Chcete vedieť viac informácií ? Pozrite sa na túto stránku.
Posledná aktualizácia: 2025-07-01 (Počet položiek: 2 206 436)
© Copyright 2025 NORMSERVIS s.r.o.