Potrebujeme váš súhlas na využitie jednotlivých dát, aby sa vám okrem iného mohli ukazovať informácie týkajúce sa vašich záujmov. Súhlas udelíte kliknutím na tlačidlo „OK“.
Standard Test Method for Determination of Acid Soluble Iron Via Atomic Absorption
Automaticky preložený názov:
Štandardná skúšobná metóda na stanovenie kyseliny rozpustné Iron Via atómovej absorpcie
NORMA vydaná dňa 1.9.2011
Označenie normy: ASTM D6647-01(2011)
Poznámka: NEPLATNÁ
Dátum vydania normy: 1.9.2011
Kód tovaru: NS-36012
Počet strán: 4
Približná hmotnosť: 12 g (0.03 libier)
Krajina: Americká technická norma
Kategória: Technické normy ASTM
Fyzikálně chemické analytické metody
Zkušební podmínky a postupy všeobecně
Keywords:
acid soluble iron, activated carbon, atomic absorption, direct aspiration: Acid-soluble iron, Atomic absorption spectrometry (AAS), ICS Number Code 19.020 (Test conditions and procedures in general), 71.040.50 (Physicochemical methods of analysis)
Significance and Use | ||||||||||||||||
In certain applications, such as acid purification, acidic food or chemical purification or decolorization, or other applications wherein iron may be leached out of the carbon, the use of acid-washed carbons will reduce or eliminate color pickup in the effluent or in the product. The acid soluble iron content is usually specified by the carbon supplier to prevent unacceptable levels of iron leach occurring. Detection limits, sensitivity, and optimum ranges will vary with the models of atomic absorption spectrophotometers. General instrumental parameters are as follows: Iron hollow cathode lamp. Wavelength: 248.3 nm. Fuel: acetylene (high purity). Oxidant: air (from compressed air line, laboratory compressor, or a cylinder of compressed airall need to be clean and dry). Type of flame: oxidizing. The following lines may also be used:
The method, as written, is intended for carbons having an acid soluble iron content of 0.0030 to 0.050 % For carbons having higher iron contents, larger dilutions or smaller aliquots may be used. The user should validate that there are no interferences from other metals in the sample matrix. To prevent erroneous results, the user should ensure that no iron instruments contact any of the sample or the solutions used in this procedure. Only glass, ceramics, or plastic should be allowed to contact the sample or solutions. |
||||||||||||||||
1. Scope | ||||||||||||||||
1.1 This test method covers the determination of the acid soluble iron content of granular and powdered activated carbons, using an atomic absorption spectroscopy method by direct aspiration. Hydrochloric acid is used to extract the iron. This test method is not directly usable to determine the total iron concentration of a sample. 1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For a specific hazard statement, see Section 8. |
||||||||||||||||
2. Referenced Documents | ||||||||||||||||
|
Poskytovanie aktuálnych informácií o legislatívnych predpisoch vyhlásených v Zbierke zákonov od roku 1945.
Aktualizácia 2x v mesiaci !
Chcete vedieť viac informácii ? Pozrite sa na túto stránku.
Posledná aktualizácia: 2024-11-04 (Počet položiek: 2 209 323)
© Copyright 2024 NORMSERVIS s.r.o.